Installation
Wiring
Thermo Scientific
Jewett Pass-Thru Blood Bank Refrigerators
11
Wiring
CAUTION
Operation under electrical conditions other than those
specified on the dataplate will cause permanent damage to the mechanism.
The data plate is located inside the refrigerator at the top of the right hand
wall.
Before connecting your refrigerator to a power source, be sure to check the
dataplate for correct voltage. Standard NEMA plugs are provided with all
units.
WARNING
For personal safety and trouble-free operation, this unit must
be properly grounded before it is used. Failure to ground the equipment
may cause personal injury or damage to the equipment. Always conform to
the National Electrical Code and local codes. Do not connect unit to already
overloaded power lines.
Always connect the equipment to a dedicated (separate) circuit. Electrical
codes require fuse or circuit breaker protection for branch circuit
conductors. Use a dedicated branch circuit breaker and time delay fuses for
#12 AWG circuits.
CAUTION
Do not use an outlet that can be turned off by a switch.
Summary of Contents for Jewett Pass-Thru BBR25D-1B
Page 4: ...Table of Contents 2 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 8: ...Inspection for Damage 6 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 14: ...Installation Wiring 12 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 32: ...Troubleshooting 32 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 34: ...Accessories 34 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 36: ...36 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...