Maintenance
Condenser Maintenance
22
Jewett Pass-Thru Blood Bank Refrigerators
Thermo Scientific
Condenser
Maintenance
CAUTION
The finned condenser can become clogged with lint and dust.
The openings between fins should be kept clean. Condensers should be cleaned
at least every six months at a minimum. In heavy traffic areas, condensers load
with heavy dirt more quickly. Failure to keep the condenser clean can result in
equipment warm-up or erratic temperatures.
WARNING
Be sure to disconnect the unit from main power before cleaning
the condenser.
CAUTION
Never clean near condensers with your fingers. Some surfaces are
sharp.
1. Disconnect the power.
2. Remove the screws on the top cage on Deck, to access the Condenser.
3. Use a vacuum cleaner with hose and brush attachments to clean the
front face of the finned surface shown in the image below.
4. Clean up any loose dust and replace the grill.
5. Reconnect the power.
Finned Surface