2-2
MaxQ 5000 Shaker
Thermo Scientific
Section 2
Specifications
Model No.
4348/4354
4349/4355
4350/4358
4351/4358
Motor
Permanent Magnet DC
Permanent Magnet DC
Solid State Brushless DC
Solid State Brushless DC
Speed
None
None
Audible with flashing LED
indicates when speed devi-
ates more than 10% of set
point
Audible with flashing LED
indicates when speed devi-
ates more than 10% of set
point
Speed Shut Off
None
None
When speed deviates 10%
of set point, unit will shut
down immediately
When speed deviates 10%
of set point, unit will shut
down immediately
Timer
None
None
Beeps twice when time has
expired. Shaking motion
stops.
Beeps twice when time has
expired. Shaking motion
stops.
Unbalanced Load
None
None
If the unit is running in an
unbalanced condition, an
alarm will sound and the
shaker will stop until the end
user corrects the condition.
The speed display will flash
“bAL” on speed panel LED.
If the unit is running in an
unbalanced condition, an
alarm will sound and the
shaker will stop until the end
user corrects the condition.
The speed display will flash
“bAL” on speed panel LED.
Optional Platform Dimensions in. (cm)
Catalog Number
L xW
30115
Universal 30” x 18” (76.2 x 45.7 cm)
Intended Use
Orbital shakers are designed to provide increased aeration in a stable
temperature environment.
Unintended Use
1) Not intended for use in Class I or II applications as defined in 21 CFR
2) Not intended for mixtures of flammable materials