In addition, all versions offer:
• Drive interrupt halts shaking action when door is opened.
• Door window.
• Foot pedal opens door automatically for hands free operation.
• All set points are retained by non-volatile memory that automatically
reactivates after power is restored.
• Visual, user adjustable over-temperature safety signal with independent
thermostat controls the heat if main controller fails.
• 1-inch (2.54 cm) triple eccentric orbital drive on 6 sets of permanently
lubricated ball bearings.
• Over temperature protection in the event of primary control failure.
• 75 lb. (34 kg) platform load capacity at safe speeds less than 400 rpm
for A-Class shakers and less than 500 rpm for E-Class shakers.
Environmental Operating Conditions
Pollution Degree 2
Installation Category II
Altitude: 2000 meters MSL (Mean Sea Level)
Relative Humidity: 20% to 80% maximum, non-condensing
Electrical Supply :120VAC or 240VAC
Voltage Tolerance: +10% of normal rated line
Temperature: 15°C to 32°C
Product Usage: Indoor use only
Caution
Shaker operation in a CO2 enriched atmosphere is not recom-
mended. Formation of carbonic acid could cause electrical failures.
s
Declaration of Conformity available on request.
2-4
MaxQ 5000 Shaker
Thermo Scientific
Section 2
Specifications