For A-Class series only:
1. Remove the 2¼-20 x ½ motor mount screws and lift the motor out.
2. Two brushes are located under plastic caps on opposite sides of the
lower part of the motor. Unscrew the plastic caps and slide the brushes
out. Replace brushes when they are worn down to 3/16" (0.476 cm) in
length.
3. Belt tension is automatically set by the location of the motor and is
not adjustable.
4. To align a pulley, loosen the motor pulley set screws, slide the pulley
up or down into alignment, then securely tighten the motor pulley set
screws.
MaxQ 5000 Shaker 6-3