Thermo Scientific Robot Stainer Microm HMS 740
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387801 - English
30
Used
: Here it is shown how often the bath
has been used.
Limit
: Is used for quality control of the
reagents. Here it can be entered how often
the bath is allowed to be used. When 90% of
the limit has been reached, a yellow frame is
shown around the bath on the screen. For
100% a red frame appears and for 110% this
frame starts to blink. To reset this number to
"0", which is shown under USED, click
RESET.
Time uncritical
: To guarantee an optimal
and efficient working with the instrument, the
first step in a staining protocol can be an
uncritical one. (See 3-4-3).
Reagent list
: To simplify the programming of
bath layouts, a list of reagents can be
created. The reagent name can be entered
under Name and can be stored via "Add to
list". For future programming the name can
easily be selected from the list.
Reagent names and descriptions can also be
deleted from the list. For this, please mark
the names, which are to be deleted, with your
finger on the touch screen and then delete
them via the field "Remove from list". For
this, mark the respective names on the touch
screen with your finger and the tip on the field
"Remove from list".