3-5-2 EDITING
STAINING PROTOCOLS
Editing an existing protocol:
Select the menu button.
Then select the submenu "PROTOCOLS".
Now select "EDIT PROTOCOLS“.
The protocol screen is shown.
Select
"LOAD".
A list of already existing protocols is shown on the
screen:
Via the arrow button select the desired protocol
and confirm with "OK".
Via the field "Stepdata", alterations can be
made.
For this, select the step which is to be altered
via the arrow buttons and enter the alteration.
In case a step is to be added, this step is placed
before the selected one. After having entered it,
acknowledge with "INSERT".
In case an already existing parameter is to be
altered in a step, it must be acknowledged with
"CHANGE".
In case a step is to be deleted, it must be marked
via the arrow buttons and acknowledged with
"DELETE".
In case a step is attached to a protocol, it must be
acknowledged with "APPEND".
When all changes are entered, click to "SAVE
AS" and enter a new protocol name.
Confirm
with
"OK".