Thermo Scientific Robot Stainer Microm HMS 740
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387801 - English
59
6-3 REPLACEMENT
WORK
6-3-1
REPLACING THE CHARCOAL FILTER
The charcoal filter should be replaced after a
maximum of 750 operating hours or every 6 - 9
months.
Note:
After 750 operating hours, a
message is shown on the screen
and asks for the replacement of the
charcoal filter.
Loosen the screws of the cover on the right
side of the bath stations.
Remove
the
cover.
Loosen and remove the screws of the air filter
container.
Remove the air filter container.
Remove the charcoal filter and replace it with
an original replacement part (cat. no. 184320).
Close the cover again and tighten the screws.
Note:
To guarantee operational security
of the instrument, only use the
charcoal filter, inserted from the
factory (cat. no. 184320).