120
11. Set the experiment parameters.
Use Experiment Setup in the collect menu to test the parameters.
(If you have set up and saved an experiment for GAO data
collection, you can set the parameters in one step by selecting or
opening the experiment.) Set Background Handling on the
Collect tab to Collect Background After Every Sample.
Note
Grazing angle measurements may require more than 1000 scans to
achieve an adequate signal-to-noise ratio, especially when band
intensities are weak.
If you need help setting parameters, find “experiment” in the
OMNIC Help system Index and go to “Using Experiment Setup.”
12. Choose collect Sample from the Collect menu to collect the
sample spectrum.
Follow the instructions that appear on the screen.
13. When you are prompted to collect the background spectrum,
position under the GAO a clean area of the reflective
substrate on which the sample is installed.
Do not adjust the Reflex aperture or the GAO.
14. While looking through the viewer (or at the video image),
finely focus the image of the substrate surface.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...