Thermo Scientific
160
Symptom
Possible Cause
Solution
No image appears in the video
pane or the viewer,
continued…
The video capture hardware
and software was incorrectly
installed or the camera driver
was incorrectly installed.
Contact technical support.
The video in the navigation
pane of the Atlµs window is
not proportional to the live
video feed in the video pane.
An optional motorized stage is
not initialized or needs to be
re-initialized.
Initialize the stage.
See
“Initializing the
stage” in the “Preparing the Microscope”
chapter if you are not familiar with the stage
initialization procedure.
Signal is weak without a
sample in beam path.
Microscope is not in correct
mode.
Select mode by setting Sample Compartment
parameter.
Find
“Sample Compartment
parameter” in OMNIC Help system Index
and go to “Specifying the sample location.”
Microscope is not focused.
Focus microscope. See “Preparing the
optics” in the “Preparing the Microscope”
chapter.
Detector is not cool.
Make sure detector dewar is filled with
liquid nitrogen. Allow enough time for
detector to cool (about 10 minutes).
See
“Cooling the detector” in “Preparing the
Microscope” chapter.
Microscope is misaligned.
Center and focus the condenser and focus the
objective.
See
“Preparing the optics” in the
“Preparing the Microscope” chapter.
When you are finished, run the Noise Test in
transmission modes
See
“Testing microscope
performance” later in this chapter if you
need help.
If the image cannot be brought into sharp
focus, performance has dropped off, or the
Noise Test fails, contact technical support.
continued…
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...