Symptom
Possible Cause
Solution
Pre-sample aperture is not
centered on reticle cross hairs.
Viewer is misaligned.
Contact technical support for assistance.
The scale in the navigation
pane is not as expected or is
zoomed in to an extreme
when the Atlµs window
opens.
An optional motorized stage is
not initialized or needs to be re-
initialized.
Initialize the stage.
See
“Initializing the
stage” in the “Preparing the Microscope”
chapter if you are not familiar with the stage
initialization procedure.
Stage does not move
predictably and the Move
Stage tools are intermittent or
do not work at all.
An optional motorized stage is
not initialized.
Initialize the stage.
See
“Initializing the
stage” in the “Preparing the Microscope”
chapter if you are not familiar with the stage
initialization procedures.
An optional motorized stage
does not move and the Move
Stage tools are not visible in
the OMNIC Atlµs Window.
The stage controller is
OFF.
Turn on the stage controller.
The microscope system was
powered on in an incorrect
sequence.
Power off the microscope, stage controller
(if installed), spectrometer and computer.
Power the system on in the following
sequence:
Microscope
Stage controller, if installed
Illuminators
Quantitative Contact Alert, if installed
Computer
Spectrometer
The stage controller cannot
communicate with the
computer.
Verify that the communications cable is
firmly seated.
The stage and/or stage
controller is malfunctioning
Replace the fast mapping stage,
Contact technical support.
continued…
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...