Continuµm User Guide 165
Symptom
Possible Cause
Solution
The optional fast mapping
stage moves erratically.
The stage calibration
parameters are not correct.
Enter the correct compensation numbers. A
label on the bottom of the stage provides the
correct compensation numbers for your
stage. Find “stage” in the OMNIC Atlµs
Help system Index and go to “Calibrating the
stage” for further information.
A detector installed does not
appear on the Bench tab of
Experiment Setup.
The microscope is turned off or
the microscope system was
powered on in an incorrect
sequence.
Power off all of the system components (the
microscope, stage controller (if installed),
spectrometer and computer).
Power the system on in the following
sequence:
Microscope
Stage controller, if installed
Illuminators
Quantitative Contact Alert, if installed
Computer
Spectrometer
The detector, computer, or
microscope is malfunctioning.
Contact technical support.
The optional stage controller,
the microscope or the computer
are malfunctioning.
Contact technical support
continued…
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...