4. Confirm the focus by looking through both eye pieces with
both eyes.
You should see one, clear image while viewing the pinhole with
both eyes.
Condenser
focus and alignment
A focused and centered condenser is important in transmission mode
for proper alignment of the infrared beam path. (In reflection mode
the condenser is not in the beam path.)
To focus the condenser:
1. Use the Sampling Mode button on the front panel to switch
the microscope into transmission sampling mode.
The transmission indicator lights whenever the microscope is in
transmission sampling mode.