The aperture automation software cannot track manual changes to
the aperture (changes made using the knobs on the right side of the
microscope). Any time you make manual changes to the aperture,
you must re-synchronize the hardware and software controls.
Notice
If you do not synchronize the aperture with the software before
loading an experiment with aperture settings or attempting to use the
software to adjust the aperture, the size and shape of the aperture
may be unpredictable.
To synchronize the physical aperture with the automated aperture
system:
1. Open OMNIC and select Transmission discrete point –
standard sample from the Experiment drop-down list box.
2. Choose Show Atlµs Window from the Atlµs menu to open the
Atlµs window.
When the Atlµs window opens, you will see navigation pane
(left side of screen) that shows the current stage location and a
video pane (right side of screen).
The colored box in the video pane represents the virtual aperture.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...