II-7
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
5 . . Wire .the .power .cable .to .the .green .screw .terminal .connector .from .the .
options .kit . . .Select .correct .terminal .for .hot .conductor .depending .on .
line .voltage . . .Refer .to .
Figure II-5
.for .terminal .connector .location .and .
Figure II-3
.for .terminal .wiring .position
.
6 . . Plug .to .terminal .connector .into .power .supply . . .Refer .to .
Figure II-3
.
7 . . Select .the .correct .fuse .from .the .fuse .kit . . .Install .by .inserting .fuse . .
in .fuse .holder .and .secure .by .twist .and .lock .method . . .Refer .to . .
Figure II-3
. . .The .fuses .are .clearly .labeled .with .the .appropriate .voltages .
for .your .system . . .Refer .to .
Figure II-3
. . .Refer .to .the .table .below .for . .
fuse .selection .
AC Voltage
Fuse Rating
115V
200mA, 250V, Fast Acting
230V
100mA, 250V Fast Acting
Figure II-5
Terminal Connector Location
Terminal Connector