III-9
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action
Marquee
Scrolling
Main Display
top line
bottom line
Notes
ALARM
ALRM
(flashing)
SEL
SCrn
Press
to set up to 3 alarms: high, low, and
an error signaling contact (rLY1, rLY2, & rLY3)
OFF
SELECT ALARM 1 HIGH
LOW OR OFF
rLY1
OFF
(Flashing)
Press
/
to select HI, LO or OFF for
rLY1
SELECT ALARM 1 HIGH
LOW OR OFF
rLY1
HI, LO or OFF
(Flashing)
Press
to set rLY1 alarm value
If OFF option is
selected the system
will proceed to rLY
00 ppb
To set HI or LO values for rLY1:
Press
to move decimal and ppm/ppb
Press
/
to set digit values
Press
to move to set next digit
ENTER VALUE
rLY1
### [concentration]
Press
to accept value and advance to next
screen for rLY2
OFF
SELECT ALARM 2 HIGH
LOW OR OFF
rLY2
OFF
(Flashing)
Press
/
to set HI, LO or OFF for rLY2
SELECT ALARM 2
HIGH LOW OR OFF
rLY2
HI, LO or OFF
(Flashing)
Press
to set rLY2 alarm value
If OFF option is
selected the system
will proceed to rLY3
ppm
To set HI or LO values for rLY2:
Press
to move decimal and ppm/ppb
Press
/
to set digit values
Press
to move to set next digit
ENTER VALUE
rLY2
### [concentration]
Press
to accept value and advance to next
screen for rLY3