III-8
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default
User Action
Marquee
Scrolling
Main Display
top line
bottom line
Notes
DKA
DKA
(flashing)
SEL
SCrn
Press
to program user customized
Double Known Calibration values
90 .0 mL
ENTER FLOW CELL VOLUME IN mL
FCEL
# # #. #
(Flashing)
This section
may not
appear,
depending
on the
system
software
revision . If
this section
does
appear,
leave the
flow cell
volume set
at 90 .0 mL
Press
/
to set flow cell
volume (mL)
ENTER FLOW CELL VOLUME IN mL
FCEL
#
# #. #
(Flashing digit will
change)
Press
to move to set each digit
Press
/
to set values of
remaining digits for flow cell volume (mL)
ENTER FLOW CELL VOLUME IN mL
FCEL
# # #.
#
(Flashing digit
will change)
Press
to accept values and advance
to next screen to set concentration for
custom standards
9 . ppm
SET STD 1 CONCENTRATION
CONC
# #. #
(Flashing)
To set value for Standard 1 concentration
(ppm or ppb):
Press
to set decimal and ppm / ppb
Press
/
to set value of digits
Press
to move to set each digit
SET STD 1 CONCENTRATION
CONC
#
# . #
(Flashing digit will
change)
Press
to accept values and advance
to next screen to set volume for Standard 1
0 .0 mL
SET STD 1 VOLUME IN mL
Std 1
#. # #
(Flashing)
To set value for Standard 1 volume (mL):
Press
/
to set value of digits
Press
to move to set each digit
SET STD 1 VOLUME IN mL
Std 1
#
#. #
(Flashing digit will
change)