IV-3
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Warning:
Be .sure .to .follow .the .air .regulation .procedure .for .rinsing .the .
flow .cell . . .Refer .to .
Chapter IV, Air Regulation
. . .Failure .to .initiate .proper .
air .settings .will .result .in .flow .cell .overfill . . .Sample .will .leak .out .the .top .of .
flow .cell .cap . . .
s
1 . . Open .calibration .shut .off .valve, .see .detail .
A
. .
. • . Open .position .shows .valve .step .in .vertical .position .facing .up .
2 . . Open .the .sample .inlet .valve, .see .detail .
B
. .
. • . Positioning .valve .with .pointed .end .facing .up .for .open .position .
3 . . Open .the .bypass .valve .by .turning .counterclockwise, .see .detail .
C
.
. • . Check .for .appropriate .sample .flow . .
. • . Adjust .pressure .regulator .if .required . . .See .step .4 .
4 . . Adjust .the .knob .on .the .regulator, . .see .detail .
E
, .so .that .ball .of .flow .
meter .reads .40 .mL/min, .see .detail .
D
. . .
. • . Pulling .the .knob .will .unlock .the .setting, .allowing .adjustment . . .
. • . . Turning .the .knob .clockwise .will .increase .flow .
. • . Pushing .the .know .back .in .will .lock .into .position . .
5 . . Turn .arrow .on .diverter .valve .to .Cal .position, .with .the .arrow .pointing .
up . . .Refer .to .
Chapter IV, Position of Valves for Calibration
. .
. • . Flow .cell .will .begin .to .fill .
6 . . Continue .to .flush .flow .cell .up .to .60 .minutes . . .Sample .will .drain .from .
siphon .tube .into .siphon .drain .to .waste .
7 . . . Proceed .with .DKA .Calibration . . .Refer .to .
Chapter IV, Before
Performing DKA Calibration
.
Rinsing the Flow Cell
50
40
30
20
10
A
D
E
C
B
Figure IV-3