IV-0
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage
User Action
Marquee
Main Display
top line
bottom line
Valve position /
Notes
30***
Turn Left Air Valve to
siphon
position. Arrow pointing down.
Turn Right Air Valve to
mix
position.
Arrow pointing down.
For correct position in
measurement mode.
Turn Diverter Valve to
sample
position. Arrow pointing left.
Turn Calibration Shut Off Valve to
measure
position. Arrow pointing
up.
Close cal port on flow cell cap.
TURN LEFT VALVE DOWN
TURN DIVERTER VALVE SO
ARROW POINTS LEFT
TURN SHUFOFF VALVE UP
CLOSE CAL PORT
PRESS ENTER
FLO On
*** Some system software
revisions will display the
combined steps shown in Stage
30. For all other systems refer
to Stages 30a, 30b, 30c, 30d, and
30e.
30a
Turn Left Air Valve to
siphon
position. Arrow pointing down
Turn Right Air Valve to
mix
position.
Arrow pointing down
For Correct position in
measurement mode
TURN LEFT AND RIGHT VALVE
TO DOWN
PRESS ENTER
PrEP rd1
30b
Press
.
30c
Turn Diverter Valve to
sample
position. Arrow pointing left
TURN DIVERTER VALVE SO
ARROW POINTS LEFT
PRESS ENTER
PrEP rd2
30d
Press
.
30e
Turn Calibration Shut Off Valve to
measure
position.
Close cal port on flow cell cap.
TURN SHUTOFF VALVE UP
CLOSE CAL PORT
PRESS ENTER
FLO On