16
Follow the steps below to set the Interior Temperture. Note: Temperature is pre-set at the factory
to -40oC. Changing the temperature setting is not reccommended.
按照以下步骤设置内部温度。
注:温度在工厂预设为
-40oC
。不建议改变温度设定。
1. Press the Set button twice.
按下
Set
(设置)
按钮
2
次
2. Press the
button or the
button to increase or decrease the Interior Temperature.
按下
按
钮或
按钮来提高或降低内部温度(
Interior Temperature
)。
Complete the steps below to adjust the Control Temperature Adjustment limits
完成下面的步骤来
改变
“
控制温度调整
”
的限值
1. Press the Set button once.
按下
Set
(设置)
按钮一次
Note: The Control Temperature Confine and the “TEMPERATURE [
℃
]” window flash together.
注:
“
控
制温度限制
”
(
Control Temperature Confine
)和
“
温度
[
℃
]”
(
TEMPERATURE
)窗口同时闪烁。
2. Press
button or
button to increase or decrease the Control Temperature Confine limits.
按下
按
钮或
按钮来提高或降低控制温度限值(
Control Temperature Confine
)。
Battery backup charge switch location and dry contact location
备用电池充电开关的位置和干式接
点位置
1. To charge the battery that is used to back up controller function only, toggle the switch located at the
back bottom of the cabinet to the “I” position
如需给用于备份控制器的电池充电,将机柜背面底部的开关
切换到
“I”
位。
2. Contacts are located at the back bottom of the cabinet.
接点位于
机柜背面底部。
Note: When setting parameters, the display screen will return to default after 8 seconds
with inactivity
注:设置参数时,
8
秒闲置后显示屏将返回到默认设置
8.2 Parameter Defaults
默认参数
The table below shows the parameter defaults for model number PLF 276
下表显示了
PLF 276
型号产品默认
的参数
Parameter
参数
Value
值
(
℃
)
Parameter
参数
Value
值
(
℃
)
Inner temperature
内部温度
-40
High alarm temperature
高温报警
-26
Control temperature confine
控制温度限值
1.1
Low alarm temperature
低温报警
-45
Adjust sensor temperature
调整传感器温度
-2.5
Table 5 Parameter Defaults PLF 276
表
5 PLF 276
默认参数