background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

6

6.3.2 Installation

Field installed systems are supplied with complete installation 
and operating instructions. If your system is factory installed, the 
freezer is shipped with a coiled length of tubing to connect the 
freezer to the bottles:

5/8 in. OD copper tubing covered with Armaflex

 insulating 

tubing for connection to the LN

supply.

Straighten the coiled tubing and connect one end to the labeled 

connection on the freezer and the other end to the supply bottle or 
building supply fitting. 

6.3.3 Operation

To activate the optional backup system, complete the following 
steps:

1. Push the Power Switch to the ON position. The digital 

temperature display window shows the current cabinet 
temperature.

2. Press and hold the Push to Set button to display the setpoint 

temperature in the digital temperature display window.

3. Use a small screwdriver to rotate the Setpoint Adjust screw 

until the correct backup operating temperature shows in the 

display window. The recommended setpoint is 10 to 15

°

above normal cabinet temperature.

Note:

The temperature shown in the display window is 
accurate to 

±

1

°

C.

The backup system can run for a minimum of eight hours on 
battery power.

WARNING!

You must push the Power Switch to the 

ON position in order to activate the backup system.

6.3.4 System Status Lights

In addition to the backup control system, the built-in backup 
system includes main control panel system status lights. These 
lights are only available when the backup system is installed.

The Supply On Indicator lights when the backup system is 
operating on battery power.

The Backup System Battery Low Indicator lights when the 
backup system battery charge is low. The built-in battery 
charger recharges the battery to full charge when power is 

restored to the system.

Note:

Since rechargeable batteries degrade over time, the 
battery should be replaced after approximately three 

years.

The Backup System Empty Indicator lights when the standby 
LN

2

 bottle is empty.

Figure 4.  Location of LN

2

 Connection Site, 

Chest Models Only

Table 1. 

Backup System Flow Rates – Cryogenic Freezers

Lbs/hr @ 75

×

Ambient Temperature 

(Flow Rate)

Center Air 
Temperature

-150

×

C

-140

×

C

-130

×

C

-120

×

C

Liquid LN

2

Chest Freezers 

7 ft

3

Empty

11.3

10.0

9.0

8.0

1/2 Full

9.7

8.6

7.6

6.8

Full

7.8

6.8

6.0

5.2

10 ft

3

Empty

15.2

13.5

11.9

10.3

1/2 Full

12.0

10.4

9.4

8.3

Full

8.9

7.8

7.0

6.1

24.75"

2.375"

LN

2

 Inlet Conn.

Copper Tubing

Factory Supplied

0.625" O.D.

MACHINE COMPARTMENT

REAR VIEW

Summary of Contents for Revco Ultima II

Page 1: ...Installation and Operation Manual Thermo Scientific Revco Ultima II Cryogenic Storage Freezers...

Page 2: ...maflex is a registered trademark of Armacell International GmBH All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries 2007 Thermo Fisher Scientific All rights rese...

Page 3: ...3 6 2 Start Up 4 6 2 1 Turning the Power On 4 6 2 2 Setting the Cabinet Temperature 4 6 2 3 Setting the Alarms 4 6 2 4 Alarm Reset Status Lights 4 6 2 5 Alarm Test 4 6 3 Backup System Optional 5 6 3 1...

Page 4: ...hermo Scientific Revco Cryogenic Storage Freezers Installation and Operation ii 8 Chart Recorders 8 8 1 Set Up and Operation 8 8 2 Power Supply 8 8 3 Changing Chart Paper 8 8 4 Calibration Adjustment...

Page 5: ...ation system is designed to maintain cryogenic temperatures with safety only when the freezer is used for storage These models may be used in an ambient environment up to 25 C 77 F WARNING This unit i...

Page 6: ...ilation no air exchange a steady refrigerant leak Under these conditions the minimum free volume requirements would be at least 86 m3 or 3006 ft3 5 2 Wiring CAUTION Connect the equipment to the correc...

Page 7: ...s pad to set the cabinet temperature refer to Section 6 2 2 on page 4 12 Cabinet Temperature Indicator This indicator lights when the temperature display window is showing the cabinet temperature refe...

Page 8: ...tch to ALARM ON The freezer is ready to operate Note If a power failure lasting over 30 seconds occurs the POWER ON light goes out and the POWER FAILURE light and digital temperature display flash sim...

Page 9: ...the cylinders in a handtruck or cart with secure chain ties for the cylinder After cylinders are connected to the equipment securely attach them with chains to a solid stationary object such as a bui...

Page 10: ...s accurate to 1 C The backup system can run for a minimum of eight hours on battery power WARNING You must push the Power Switch to the ON position in order to activate the backup system 6 3 4 System...

Page 11: ...r standing in the bottom of the cabinet 6 If there is freezer odor wash the interior with a solution of baking soda and warm water Clean the exterior with any common household cleaning wax 7 Close the...

Page 12: ...pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the reference mark a small groove on the left side of...

Page 13: ...entifizieren die unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli...

Page 14: ...can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number...

Page 15: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 39125H51 Rev D...

Reviews: