background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

7

7 Maintenance and Troubleshooting

WARNING!

Unauthorized repair of your freezer will 

invalidate your warranty. Contact Technical Service at 
1-800-438-4851 for additional information.

CAUTION!

Maintenance should only be performed 

by trained personnel.

7.1

Condenser Maintenance

7.1.1 Cleaning the Condenser

Clean the condenser at least every six months; more often if the 

laboratory area is extremely dust prone. 

To clean the condenser, complete the following steps:

1. Pull the grill open.

2. Remove the filter. Check the fans. If a fan is not operating, 

contact an Authorized Service Company immediately.

3. Vacuum the condenser.

4. Replace the filter and close the grill.

7.1.2 Cleaning the Condenser Filter

Clean the condenser filter every two or three months.

1. Pull the grill open.
2. Remove the filter.
3. Shake the filter to remove loose dust, rinse the filter in clean 

water, shake the excess water from the filter, and replace the 
filter.

4. Close the grill.

7.2

Gasket Maintenance

Periodically check the gaskets around the door or lid for 
punctures or tears. Leaks are indicated by a streak of frost which 

forms at the point of gasket failure. Make sure that the cabinet is 
level (refer to Section 5.3 on page 2 for leveling information).

Keep the lid gasket clean and frost free by wiping gently with a 

soft cloth.

7.3

Defrosting the Freezer

Defrost the freezer once or year or whenever the ice buildup 
exceeds 3/8”.

To defrost, complete the following steps:

1. Remove all products and place in another cabinet.
2. Turn off the freezer.

3. Open the outer door and all inner doors.
4. Let the freezer stand with the door open for at least 24 hours. 

This allows both the interior and foamed refrigerant system 
to warm to room temperature.

5. Dispose of the ice and wipe out any water standing in the 

bottom of the cabinet.

6. If there is freezer odor, wash the interior with a solution of 

baking soda and warm water. Clean the exterior with any 
common household cleaning wax.

7. Close the doors, restart the freezer and reload, following the 

instructions in Section 3.1 on page 1.

7.4

Alarm Battery Maintenance

Have a technician check the condition of the alarm battery at 
least once a year.

To replace the alarm battery, complete the following steps:

1. Remove the front grill. The alarm battery is located directly 

behind the grill. The terminals are the “push on” type.

2. Grasp the terminal with pliers and work it gently back and 

forth while pulling it off. The fittings are tight.

3. Remove the battery and put the new battery in place.

Note:

You may have to cut a strip of silicone rubber in order to 
remove the battery.

4. Connect the battery terminals and replace the front grill.

7.5

Frame Heater Switch

Cryogenic chest freezers (-140º and -150º C) have a frame heater 
switch which can be used to prevent or remove condensation 
near the freezer compartment lid. The switch is located in back of 

the unit near the top center, to the right of the junction box. 
Turning the switch to the up position heats up the frame around 
the lid. 

Heating the frame has no effect on interior cabinet temperature. 
You can turn on the switch whenever there is condensation near 
the lid.

7.6

Troubleshooting Procedures

Table 2. 

Troubleshooting Procedures

Problem

Solution

Check Fuse 
indicator is on.

Surge protector is blown. Call an Electronics 
Technician to check the equipment.

Clean Filter 
indicator is on.

Condenser filter is dirty. Refer to Section 7.1.2.

Alarm Battery Low 
indicator is on.

Alarm battery is low on charge. Have a 
technician check the alarm battery. Refer to 
Section 7.4.

Voltage Boost 
indicator is on.

Low incoming line voltage to the freezer. Call an 
electronics technician to check the power supply.

No display, no fans 
running.

No power to the unit. Call a qualified electrical 
service technician to check the service breaker 
to the outlet.

Summary of Contents for Revco Ultima II

Page 1: ...Installation and Operation Manual Thermo Scientific Revco Ultima II Cryogenic Storage Freezers...

Page 2: ...maflex is a registered trademark of Armacell International GmBH All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries 2007 Thermo Fisher Scientific All rights rese...

Page 3: ...3 6 2 Start Up 4 6 2 1 Turning the Power On 4 6 2 2 Setting the Cabinet Temperature 4 6 2 3 Setting the Alarms 4 6 2 4 Alarm Reset Status Lights 4 6 2 5 Alarm Test 4 6 3 Backup System Optional 5 6 3 1...

Page 4: ...hermo Scientific Revco Cryogenic Storage Freezers Installation and Operation ii 8 Chart Recorders 8 8 1 Set Up and Operation 8 8 2 Power Supply 8 8 3 Changing Chart Paper 8 8 4 Calibration Adjustment...

Page 5: ...ation system is designed to maintain cryogenic temperatures with safety only when the freezer is used for storage These models may be used in an ambient environment up to 25 C 77 F WARNING This unit i...

Page 6: ...ilation no air exchange a steady refrigerant leak Under these conditions the minimum free volume requirements would be at least 86 m3 or 3006 ft3 5 2 Wiring CAUTION Connect the equipment to the correc...

Page 7: ...s pad to set the cabinet temperature refer to Section 6 2 2 on page 4 12 Cabinet Temperature Indicator This indicator lights when the temperature display window is showing the cabinet temperature refe...

Page 8: ...tch to ALARM ON The freezer is ready to operate Note If a power failure lasting over 30 seconds occurs the POWER ON light goes out and the POWER FAILURE light and digital temperature display flash sim...

Page 9: ...the cylinders in a handtruck or cart with secure chain ties for the cylinder After cylinders are connected to the equipment securely attach them with chains to a solid stationary object such as a bui...

Page 10: ...s accurate to 1 C The backup system can run for a minimum of eight hours on battery power WARNING You must push the Power Switch to the ON position in order to activate the backup system 6 3 4 System...

Page 11: ...r standing in the bottom of the cabinet 6 If there is freezer odor wash the interior with a solution of baking soda and warm water Clean the exterior with any common household cleaning wax 7 Close the...

Page 12: ...pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the reference mark a small groove on the left side of...

Page 13: ...entifizieren die unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli...

Page 14: ...can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number...

Page 15: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 39125H51 Rev D...

Reviews: