background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

8

8 Chart Recorders 

Panel-mounted six-inch seven-day recorders are available as 
options for all freezer models.

8.1 Set Up and Operation

To prepare the recorder to function properly, complete the 

following steps:

1. Open the recorder door to access the recorder.
2. Connect the nine volt DC battery located at the recorder’s 

upper right corner. This battery provides backup power.

3. Install clean chart paper (refer to Section 8.3 below).
4. Remove the plastic cap from the pen stylus and close the 

recorder door.

Recorder operation begins when the system is powered on.The 
recorder may not respond until the system reaches temperatures 
within the recorder’s range. 

Figure 5.  Chart Recorder

Figure 6.  Chart Buttons

8.2 Power Supply

The recorder normally uses AC power when the system is 
operating. If AC power fails, the LED indicator flashes to alert 
you to a power failure. The recorder continues sensing cabinet 
temperature and the chart continues turning for approximately 

24 hours with back-up power provided by the nine-volt battery. 

The LED indicator glows continuously when main power is 
functioning and the battery is charged.

When the battery is low, the LED flashes to indicate that the 
battery needs to be changed.

8.3 Changing Chart Paper

To change the chart paper, complete the following steps:

1. Locate the pressure sensitive buttons at the front, upper left 

of the recorder panel.

2. Press and hold the Change Chart button (#3) for one second. 

The pen will move off the scale.

3. Unscrew the center nut, remove the old chart paper, and 

install new chart paper. Carefully align the day and time with 

the reference mark (a small groove on the left side of the 
recorder panel).

4. Replace the center nut and hand tighten. Press the Change 

Chart button again to resume temperature recording.

8.4 Calibration Adjustment

This recorder has been accurately calibrated at the factory and 
retains calibration even during power interruptions. If required, 

however, adjustments can be made as follows:

1. Run the unit continuously at the control setpoint temperature. 

Continue steady operation for at least two hours to provide 

adequate time for recorder response.

2. Measure cabinet center temperature with a calibrated 

temperature monitor.

3. Compare the recorder temperature to the measured cabinet 

temperature. If necessary, adjust recorder by pressing the left 
(#1) and right (#2) chart buttons.

Note:

The stylus does not begin to move until the button is held 

for five seconds.

If you find that the recorder frequently needs to be calibrated 
after paper changes, the program may have changed. To confirm 

that the program is correct:

1. Press button #3 and allow the pin to move off the scale.
2. Press #1 for 10 seconds and release.

3. Confirm that the green LED flashes 3 times followed by a 

pause. If the number of blinks is greater than or less than 3, 
adjust the number using the left (#1) and right (#2) chart 

buttons.

4. When the green LED blinks 3 times to confirm the resetting, 

press the #3 button. The pin arm should then return to 
indicate the current temperature. Repeat Steps 1-3 if 

necessary.

9 Volt Battery

I

mprinting

Stylus

Hub-Nut and
Retaining Wire

Chart

Chart Buttons

Reference 
Mark

1

2

3

CHANGE CHART

Summary of Contents for Revco Ultima II

Page 1: ...Installation and Operation Manual Thermo Scientific Revco Ultima II Cryogenic Storage Freezers...

Page 2: ...maflex is a registered trademark of Armacell International GmBH All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries 2007 Thermo Fisher Scientific All rights rese...

Page 3: ...3 6 2 Start Up 4 6 2 1 Turning the Power On 4 6 2 2 Setting the Cabinet Temperature 4 6 2 3 Setting the Alarms 4 6 2 4 Alarm Reset Status Lights 4 6 2 5 Alarm Test 4 6 3 Backup System Optional 5 6 3 1...

Page 4: ...hermo Scientific Revco Cryogenic Storage Freezers Installation and Operation ii 8 Chart Recorders 8 8 1 Set Up and Operation 8 8 2 Power Supply 8 8 3 Changing Chart Paper 8 8 4 Calibration Adjustment...

Page 5: ...ation system is designed to maintain cryogenic temperatures with safety only when the freezer is used for storage These models may be used in an ambient environment up to 25 C 77 F WARNING This unit i...

Page 6: ...ilation no air exchange a steady refrigerant leak Under these conditions the minimum free volume requirements would be at least 86 m3 or 3006 ft3 5 2 Wiring CAUTION Connect the equipment to the correc...

Page 7: ...s pad to set the cabinet temperature refer to Section 6 2 2 on page 4 12 Cabinet Temperature Indicator This indicator lights when the temperature display window is showing the cabinet temperature refe...

Page 8: ...tch to ALARM ON The freezer is ready to operate Note If a power failure lasting over 30 seconds occurs the POWER ON light goes out and the POWER FAILURE light and digital temperature display flash sim...

Page 9: ...the cylinders in a handtruck or cart with secure chain ties for the cylinder After cylinders are connected to the equipment securely attach them with chains to a solid stationary object such as a bui...

Page 10: ...s accurate to 1 C The backup system can run for a minimum of eight hours on battery power WARNING You must push the Power Switch to the ON position in order to activate the backup system 6 3 4 System...

Page 11: ...r standing in the bottom of the cabinet 6 If there is freezer odor wash the interior with a solution of baking soda and warm water Clean the exterior with any common household cleaning wax 7 Close the...

Page 12: ...pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the reference mark a small groove on the left side of...

Page 13: ...entifizieren die unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli...

Page 14: ...can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number...

Page 15: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 39125H51 Rev D...

Reviews: