Installation and Operation

1

1 About Cryogenic Models

The freezers described in this manual are 

cryogenic

 models. 

These models are designed for optimum performance at -140ºC 
or -150ºC, and it is strongly recommended that you use these 

models at these operating temperatures.

Be sure to read the maintenance instructions in Section 7.

2 Safety Precautions

In this manual and on labels attached to this product, the words 
WARNING and CAUTION mean the following:

WARNING: a potentially hazardous situation which, if not 
avoided, could result in serious injury or death.

CAUTION: a potentially hazardous situation which, if not 
avoided, may result in minor or moderate injury or damage to 
the equipment.

Before installing, using or maintaining this product, please be sure 
to read this manual and product warning labels carefully. Failure to 
follow these instructions may cause this product to malfunction, 
which could result in injury or damage.

Below are important safety precautions that apply to this product:

Use this product only in the way described in the product 
literature and in this manual. Before using it, verify that this 
product is suitable for its intended use. 

Do not modify system components, especially the controller. 
Use OEM exact replacement equipment or parts. Before use, 
confirm that the product has not been altered in any way.

Your unit must be properly grounded in conformity with 
national and local electrical codes. Never connect the unit to 
overloaded power sources.

Disconnect the unit from all power sources before cleaning, 
troubleshooting, or performing other maintenance on the 
product or its controls.

3 General Recommendations

The refrigeration system is designed to maintain cryogenic 
temperatures with safety 

only

 when the freezer is used for 

storage. These models may be used in an ambient environment 
up to +25ºC (77ºF). 

WARNING!

This unit is not a “rapid-freeze” device. 

Freezing large quantities of liquid, or high-water 
content items, will temporarily increase the chamber 
temperature and will cause the compressors to 
operate for a prolonged time period. 

Avoid opening the door for extended time periods since chamber 

temperature air will escape rapidly. Room air, which is higher in 
humidity, replacing chamber air may cause frost to develop in the 
chamber more rapidly.

3.1 Initial Loading

After reading and completing the Safety Considerations, 
Pre-Installation, Installation, and Operation sections of this 
manual, turn the Key Switch to the POWER ON position and 
adjust the temperature setpoint if necessary.

The setpoint should be no warmer than -50ºC. 

Allow the freezer 

to operate at the desired temperature for a minimum of 12 
hours before loading.

CAUTION!

IFailure to follow these procedures or 

overloading the unit may cause undue stress on the 
compressors or jeopardize user product safety.

4 Pre-Installation

4.1 Unpacking

At delivery, examine the exterior for physical damage while the 

carrier’s representative is present. If exterior damage is 
present, carefully unpack and inspect the unit and all 
accessories for damage.

If there is no exterior damage, unpack and inspect the 
equipment within five days of delivery. If you find any damage, 
keep the packing materials and immediately report the damage 

to the carrier. 

Do not return goods without written 

authorization

. When submitting a claim for shipping damage, 

request that the carrier inspect the shipping container and 

equipment.

CAUTION!

Do not discard the sublids from 

chest-style units.

 The sublids are necessary to 

maintain correct temperature, moisture control, and 
economy of operation. 

Summary of Contents for Revco Ultima II

Page 1: ...Installation and Operation Manual Thermo Scientific Revco Ultima II Cryogenic Storage Freezers...

Page 2: ...maflex is a registered trademark of Armacell International GmBH All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries 2007 Thermo Fisher Scientific All rights rese...

Page 3: ...3 6 2 Start Up 4 6 2 1 Turning the Power On 4 6 2 2 Setting the Cabinet Temperature 4 6 2 3 Setting the Alarms 4 6 2 4 Alarm Reset Status Lights 4 6 2 5 Alarm Test 4 6 3 Backup System Optional 5 6 3 1...

Page 4: ...hermo Scientific Revco Cryogenic Storage Freezers Installation and Operation ii 8 Chart Recorders 8 8 1 Set Up and Operation 8 8 2 Power Supply 8 8 3 Changing Chart Paper 8 8 4 Calibration Adjustment...

Page 5: ...ation system is designed to maintain cryogenic temperatures with safety only when the freezer is used for storage These models may be used in an ambient environment up to 25 C 77 F WARNING This unit i...

Page 6: ...ilation no air exchange a steady refrigerant leak Under these conditions the minimum free volume requirements would be at least 86 m3 or 3006 ft3 5 2 Wiring CAUTION Connect the equipment to the correc...

Page 7: ...s pad to set the cabinet temperature refer to Section 6 2 2 on page 4 12 Cabinet Temperature Indicator This indicator lights when the temperature display window is showing the cabinet temperature refe...

Page 8: ...tch to ALARM ON The freezer is ready to operate Note If a power failure lasting over 30 seconds occurs the POWER ON light goes out and the POWER FAILURE light and digital temperature display flash sim...

Page 9: ...the cylinders in a handtruck or cart with secure chain ties for the cylinder After cylinders are connected to the equipment securely attach them with chains to a solid stationary object such as a bui...

Page 10: ...s accurate to 1 C The backup system can run for a minimum of eight hours on battery power WARNING You must push the Power Switch to the ON position in order to activate the backup system 6 3 4 System...

Page 11: ...r standing in the bottom of the cabinet 6 If there is freezer odor wash the interior with a solution of baking soda and warm water Clean the exterior with any common household cleaning wax 7 Close the...

Page 12: ...pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the reference mark a small groove on the left side of...

Page 13: ...entifizieren die unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli...

Page 14: ...can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number...

Page 15: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 39125H51 Rev D...

Reviews: