Smart-Vue Pro Duo / Quatro User Guide
Technical Specifications | 11
Technical Specifications
Characteristics
General
•
Color LCD display: 2.4 inches
•
Touch screen: Resistive
•
Number of sensor channels (simultaneous data logging
sessions): Smart-Vue Pro Duo: 2 channels, Smart-Vue
Pro Quatro: 4 channels.
•
Number of sensors:
Smart-Vue Pro Duo/Quatro: One physical plug (supports
1 single, dual or triple sensor).
Smart-Vue Pro Quatro: Two physical plugs (supports 2
single, dual or triple sensors).
Note:
It is important to distinguish between the number of
sensors assigned to a given data logger and the number of
simultaneous data logging sessions you can have.
•
Audible alarms: Buzzer
•
Power Supply: Batteries or micro-USB (5 V)
Note:
The AC power adapter is provided separately and
has: 5V DC/1.2A. Batteries must be replaced by authorized
technicians only.
Wireless Technologies
•
LoRaWAN® wireless connectivity for data transmission:
•
Range: approximately 16 km (about 10 miles) in line-
of-sight.
•
Frequency (worldwide): 868 MHz (EU) or 915 MHz
(USA).
•
Max power output: 14 dB (EU) or 20 dB (USA).
•
Bluetooth Low Energy (BLE, also referred to as Bluetooth
Smart) technology for use with Smart Connect mobile
application:
•
Range: Up to about 50 meters (160 ft.) in
line-of-sight.
•
Frequency: 2.4 GHz (worldwide).
Monitoring
•
Read interval with Smart Connect operation: 15sec to
12 hours.
•
Read interval with LoRaWAN operation: 1 minute to 12
hours (depending on the transfer interval configured in
the Smart-Vue Pro web application).
•
Data storage:
•
Unlimited on Smart-Vue Pro web platform.
•
4,000 readings (per sensor channel) stored in internal
data logger memory.
•
1, 2 or 4 sensor tiles displayed simultaneously.
•
Color LED indicator for alarm status or bluetooth
communication.
Operating and Storage
Conditions
•
Indoor use only in non-harsh environments with mounting
height under 2 meters (about 6.5 feet) from the floor.
•
Designed for altitudes up to 6,500 feet (2,000 meters).
•
Screen operating range: 0 to + 50 °C.
•
Data logger operating range: 0 °C to + 50 °C and 0 to
90% RH (non-condensing).
•
Data logger storage conditions: -10°C to + 60 °C
(14 °F and 140 °F); 0 to 99.99% relative humidity
non-condensing.
•
Pollution degree: 2 (normally only non-conductive
pollution is supported; temporary conductivity caused by
condensation is to be expected).
Casing & Dimensions
•
Product protection: Designed for indoor use.
•
Casing: ABS plastic
•
Weight: 180 g (6.4 oz.) with batteries
•
Dimensions: 100.8 x 110.8 x 296 mm (4 x 4.4 x 1.1 in.)
•
Mounting: Magnet or screw mount (Optional locking with
padlock).