Smart2Pure Pro Ultrapure water system
Installation | 21
•
Connect 6 mm tubing from assembly kit to the Tank Back
Flow connector on Smart2Pure Pro (to be connected to
tank)
•
Connect tank level and pump control cable (from tank) to
the Level Control connector on the Smart2Pure Pro
Figure 18. Tank Flow connection
4. Connect 6 mm tubing from Tank Flow connector on
Smart2Pure Pro, to the Tank Flow connector on Tank
using straight reduction connector (part no. - 14.0318)
and 8 mm tube from kit (refer
).
Connect 6 mm tubing from Tank Back Flow connector on
Smart2Pure Pro, to the Tank Back Flow connector on
Tank
5. Connect ¼” tubing from concentrate connector on
Smart2Pure Pro to atmospherically vented drain
6. Connect optional remote hand dispenser to the
Dispenser Flow and Dispenser Back Flow connector on
Smart2Pure Pro
7. Connect optional remote hand dispenser to the ¾” NPT
pressurized outlet connector on bottom right of pure
water tank
Alternative Feed Water
Connections
If pretreatment other than the part no. 50157886 is used, then
use the R¾” feed-water connecting kit (refer
) or
Female Adaptor “BSP Thread x” Tube OD #50157861,
supplied with the system to connect to alternative pretreated
feed water supply. Assemble feed water connector kit as
shown in
. Insert Smart2pure Pro ¼” feed water
tubing into the fitting and then connect the union nut to the
supply.
Figure 19. Feed-water Smart2Pure Pro connector
Routing Tubes to Drain
Note:
Shorten the rinse water and concentrate outlet tube to
the required length and route it to an atmospherically vented
drain. The tubes that run from the Water Purification system
and the external tank to the drain must be routed with a
downward slope and without any kinks or restrictions, as this
would result in backing up of the draining water.
If a standard drain siphon is in place, the ends of the tubes
must be located at least 20 mm / 0.75 inch above the drain.
Attach the tubes in such a manner that they remain in their
position.
Figure 20. Routing drain pipes.
Installation of Ultrapure
Cartridge and RO Cartridge
Note:
•
After Installation all tubing must be checked for their
correct position on the systems panel and their is no
leakage or blockage occurs after opening the feed-water
supply.
•
To avoid tripping, ensure that the tubing and wires do not
lay over the floor.
1. Remove the cover from the Smart2Pure Pro by pressing
the snapper and pull the cover off towards the front.
Connect 6 mm tube
from Smart2Pure Pro
8 mm tube
(Cut approx. 3” to 4”)
Part No. - 14.0318
Overflow tube
Rinse water tube
Minimum distance
20 mm / 0.75 inch
Drain