36 | Maintenance
Smart2Pure Pro Ultrapure water system
3. Press on the two quick connectors at the inlet and outlet
of the ultrapure cartridge and remove the used cartridge
from the system.
Figure 39. Removing the ultrapure quick
connectors
4. We recommend performing a disinfection when an
existing ultrapure cartridge is replaced.
Note:
Refer to section
for an explanation of
how to perform disinfection.
5. Locate the new ultrapure cartridge and insert the
cartridge into the system.
6. Insert the quick connectors into the connecting points on
the ultrapure cartridge. When you hear an audible click
you can be sure that the quick connectors have been
inserted correctly.
Figure 40. Installing the quick connectors
Note:
The quick-connectors are attached to the system
in such a manner so it is to prevent any confusion
(switching).
7. Put the cover for the ultrapure cartridge back in place
and switch the system ON (connect the main power
supply).
Note:
Discharge the first 5 liters and check the production of
ultrapure water by dispensing water from the tap. If there is no
ultrapure water flow, check level of water in storage tank and
repeat the steps of replacing the ultrapure cartridge.
Replacing the RO
Cartridge
Note:
Replace the RO
cartridge
when the maximum limit for
the feed-water has been exceeded, when you cannot extract
any water from the sampling tap or when the replacement
interval is due.
1. Switch the system OFF (unplug main power supply).
2. Remove the cover over the RO
cartridge
by pressing the
snapper and pull the cover forward.
Figure 41. Remove the cover
3. Press on the three quick connectors on the cover of the
RO
cartridge
and remove the used cartridge from the
system.
Figure 42. Removing the RO quick connectors
4. Locate the new RO cartridge and insert the cartridge into
the system.
5. Insert the quick connectors into the connecting points on
the new RO
cartridge
. When you hear an audible click
you can be sure that the quick-action fasteners have
Quick
Ultrapure
Connectors
Cartridge
Ultrapure
Cartridge
Snapper
Cover
Quick Connectors
RO
cartridge
Top of
Bottom of
RO cartridge
RO cartridge