Thermo Fisher Scientific ULT Freezers
Appendix B: Modbus ASCII Parameter Table | 33
S.No
Parameter
Function
code
Address in
hexadecimal
size
Modbus
command
Relay
Enclosure
response
Data
Data
type
Remarks
10
TC3
0x03
4D0
2
3a 31 39
30 33
30
34 44 30
30 30 30
32 30 45
0d 0a
3A 31 39
30 33 30
34
34 31
42 42 30
43 41 43
32 34 0D
0A
0x41
BB0
CAC
Float
Convert the Float to
decimal with below steps:
1.Open the link
https://www.h-schmidt.net/
FloatConverter/IEEE754.html
2.Paste the data in
“Hexadecimal Representation”
and press Enter.
3.The temp value is shown in
“Decimal representation”
11
TC4
0x03
4D4
2
3a 31 39
30 33
30
34 44 34
30 30 30
32 30 41
0d 0a
3A 31 39
30 33 30
34
34 31
45 35 34
31 36 45
43 37 0D
0A
0x41
E581
6E
Float
Convert the Float to
decimal with below steps:
1.Open the link
https://www.h-schmidt.net/
FloatConverter/IEEE754.html
2.Paste the data in
“Hexadecimal Representation”
and press Enter.
3.The temp value is shown in
“Decimal representation”
12
TC5
0x03
4D8
2
3a 31 39
30 33
30
34 44 38
30 30 30
32 30 36
0d 0a
3A 31 39
30 33 30
34
43 32
46 45 30
30 30 30
31 43 0D
0A
0xC2
FE00
00
Float
Convert the Float to
decimal with below steps:
1.Open the link
https://www.h-schmidt.net/
FloatConverter/IEEE754.html
2.Paste the data in
“Hexadecimal Representation”
and press Enter.
3.The temp value is shown in
“Decimal representation”
13
TC6
0x03
4DC
2
3a 31 39
30 33
30
34 44 43
30 30 30
32 30 32
0d 0a
3A 31 39
30 33 30
34
43 32
46 45 30
30 30 30
31 43 0D
0A
0xC2
FE00
00
Float
Convert the Float to
decimal with below steps:
1.Open the link
https://www.h-schmidt.net/
FloatConverter/IEEE754.html
2.Paste the data in
“Hexadecimal Representation”
and press Enter.
3.The temp value is shown
in “Decimal representation”