34 | Appendix B: Modbus ASCII Parameter Table
Thermo Fisher Scientific ULT Freezers
14
TC7
0x03
2E0
2
3a 31 39
30 33
30
34 45 30
30 30 30
32 46 45
0d 0a
3A 31 39
30 33 30
34
43 32
46 45 30
30 30 30
31 43 0D
0A
0xC2
FE00
00
Float
Convert the Float to
decimal with below steps:
1.Open the link
https://www.h-schmidt.net/
FloatConverter/IEEE754.html
2.Paste the data in
“Hexadecimal Representation”
and press Enter.
3.The temp value is shown in
“Decimal representation”
15
TC8
0x03
4E4
2
3a 31 39
30 33
30
34 45 34
30 30 30
32 46 41
0d 0a
3A 31 39
30 33 30
34
34 31
42 43 43
44 39 46
37 33 0D
0A
0x41
BCC
D9F
Float
Convert the Float to
decimal with below steps:
1.Open the link
https://www.h-schmidt.net/
FloatConverter/IEEE754.html
2.Paste the data in
“Hexadecimal Representation”
and press Enter.
3.The temp value is shown in
“Decimal representation”
16
TC9
0x03
4E8
2
3a 31 39
30 33
30
34 45 38
30 30 30
32 46 36
0d 0a
3A 31 39
30 33 30
34 34 31
42 45 32
34 41 31
31 34 0D
0A
0x41
BE28
A1
Float
Convert the Float to
decimal with below steps:
1.Open the link
https://www.h-schmidt.net/
FloatConverter/IEEE754.html
2.Paste the data in
“Hexadecimal Representation”
and press Enter.
3.The temp value is shown in
“Decimal representation”
17
TC10
0x03
4EC
2
3a 31 39
30 33
30
34 45 43
30 30 30
32 46 32
0d 0a
3A 31 39
30 33 30
34
34 31
42 45 32
43 37 45
33 33 0D
0A
0x41
BE2
C7E
Float
Convert the Float to
decimal with below steps:
1.Open the link
https://www.h-schmidt.net/
FloatConverter/IEEE754.html
2.Paste the data in
“Hexadecimal Representation”
and press Enter.
3.The temp value is shown in
“Decimal representation”
S.No
Parameter
Function
code
Address in
hexadecimal
size
Modbus
command
Relay
Enclosure
response
Data
Data
type
Remarks