12 | Getting to Know Your Product
1.8 cu.ft. Countertop Refrigerator
Product Exterior – Side
Pass Through Port
An additional pass through port is located at the access
panel. The panel does not need to be removed to utilize the
pass through port.
Battery Monitoring and
Maintenance
The battery compartment is located on the lower rear of the
left side panel of the unit. This battery only backs up the data
monitoring system. It does not run the cooling system. The
monitoring system is designed to display an alert when the
battery drops below 30% remaining capacity.
The battery serves to support continuous temperature
monitoring during power outages and will need to be replaced
periodically.
Activating the Data Monitoring
Backup Battery
The included battery is shipped with a protective tab so that
its power doesn’t drain. To activate it you must remove the
red tab from the access panel.
Replacing the Battery
1. Remove the 2 screws securing the pass through port and
access panel.
2. Remove the panel.
3. Remove and replace with a CR123A or IEC16340 type
battery.
4. Replace the panel and tighten the 2 screws.
Note:
We recommend that you keep spare batteries on-
hand. This battery will allow you to continue to monitor
internal temperature even in the event of a power outage.
Product Exterior – Rear
Fans
The TSG205 refrigeration system is a solid-state system with
two external fans with moving parts. The functioning of the
fans is integral to heat dissipation. The fans are mounted on
When replacing the battery, replace only with a
CR123A or IEC16340 type battery.
This battery will not power the refrigeration. You
must ensure uninterrupted power to the
refrigerator and monitoring system at all times.