22 | Settings
1.8 cu.ft. Countertop Refrigerator
Setting Retention Values
In the event of a reset or power outage, the following table illustrates the various product Setting Retention Values:
Setting
Default Value on
Initial Power-Up
Setting retained
After Pushbutton
Reset?
Setting Retained
After AC Power
cycle with Battery?
Door Ajar Alarm On/Off
ON
Yes
Yes
Door Ajar Alarm Interval Setting
1-min
Yes
Yes
Temp Set Point
5.0
°C
Yes
Yes
Temp Upper Alarm Threshold
8.0
°C
Yes
Yes
Temp Lower Alarm Threshold
2.0
°C
Yes
Yes
Time Display Setting
°C
Yes
Yes
Temp Resolution Setting
0.1
°C
Yes
Yes
Temp Display (Control/Glycol)
Glycol
Yes
Yes
Memory Alarm On/Off
Enabled
Yes
Yes
Memory Capacity Alarm Setting
25%
Yes
Yes
Battery Alarm On/Off
Enabled
Yes
Yes
Battery Alarm % Remaining Setting
30%
Yes
Yes
Power Out Alarm On/Off
ON
Yes
Yes
Time and Date
01/01/2018;12:00:00pm
Yes, after manually set
Yes, after manually set
Data Log Interval
15 Minutes
Yes
Yes
Glycol Temp Calibration Value
Func 2 Calibration
Yes
Yes
Control Temp Calibration Value
Func 2 Calibration
Yes
Yes