Summary of Contents for TSG1500F

Page 1: ...l Purpose Refrigerator and Freezers Operating and Maintenance Instructions 332019H01 Revision A 03 05 2021 Please visit www thermofisher com user manuals for manuals in French Italian and Spanish lang...

Page 2: ...med by qualified service personnel Material in this manual is for informational purposes only The contents and the product it describes are subject to change without notice Thermo Fisher Scientific ma...

Page 3: ...port 3 Safety 4 General Safety Rules 4 Safety and Accident Prevention 4 Contradictions 4 Warning on the Refrigerant Gas 5 Product Description 6 Technical Description 6 Intended Use 6 Transport and Han...

Page 4: ...e Removal 17 Cleaning the Unit 17 Cleaning the Condenser 17 Condensate Water Draining 18 Replacing Buffer Batteries 18 Disposal 19 Labeling 20 Machine Data Plates 20 Other Labeling 21 Consumable Mater...

Page 5: ...nd Freezers Models 1 Models The following table shows the models covered in this Operating and Maintenance Manual Table 1 Applicable Models Models Temperature TSG400R 2 to 10 C TSG700R TSG1500R TSG400...

Page 6: ...will become immediately void in the event of use or maintenance of equipment not in compliance with specifications by the manufacturer ThermoFisher Scientific Provide all the information required rega...

Page 7: ...refully and make reference to it Routine maintenance worker Qualified technician able to carry out routine maintenance on the machine following the instructions in this manual Special maintenance work...

Page 8: ...rs Within the limits permitted by their functions accessible parts of the machine have no sharp corners sharp edges or rough surfaces that could cause injury Moving parts All components with the possi...

Page 9: ...gas contained in the circuit is declared on the silver data plate attached inside the cabinet Do not use the equipment if damaged Make sure that air grids are always free to guarantee a certain venti...

Page 10: ...asily replaceable magnetic seals for perfect tightness Measures were taken during design and construction to obtain a machine that meets specific safety requirements such as internal rounded corners c...

Page 11: ...If the machine is embedded in any type of furniture a correct air flow of the condensing unit compressor fan motors must be always ensured The warranty will immediately expire if this is not guarante...

Page 12: ...er Scientific devices are equipped with 2 safety fuses phase and neutral with integral protection against electric shocks short circuits and over currents and according to the standards for laboratory...

Page 13: ...mpartment Observe the following instructions Do not place material over the label indicating the maximum permitted loading level if present Store material leaving a distance of at least 6 cm from side...

Page 14: ...otor compartment air vents Load stored material gradually at ambient temperature to grant proper refrigeration Store material on shelves or drawers Do not place products directly on the bottom or agai...

Page 15: ...tch on the equipment for the first time follow these below instructions Table 2 Keyboard function UI Command To confirm the selection Value increase button menu scrolling and DOCTOR VIEW function To e...

Page 16: ...etpointand temperature Date and time Relative humidity and temperature Table 4 Action Strings String Operation in Progress Pause The compressor is OFF waiting for next cooling cycle Cool The compresso...

Page 17: ...anual defrosting This function is enabled by following the procedure below Press the button and by using the buttons or stops at the string Confirm by pressing the button The display will show a messa...

Page 18: ...ly in use will be requested which then you will be able to edit if any password has been entered nothing is requested Select the password in use using the buttons or and confirm with If entered correc...

Page 19: ...side the chamber LOW TEMP Low temperature inside the chamber POWER FAIL HT High temperature due to power failure MAINS FAILURE Power failure DOOR Open door PROBE S1 Faulty thermostat control probe cal...

Page 20: ...e than 2 minutes the display will show the message MAINS FAILURE alternating with the frame of the selected mode The buzzer will activate and the display will flash for 5 minutes and then start again...

Page 21: ...a year or at different intervals depending on the type products stored Cleaning the Condenser Failure to clean the condenser as well as temperature being too high in the environment in which it is in...

Page 22: ...estore the electrical current and switch back on the device Figure 18 Condenser Position Condensate Water Draining Defrosting causes the formation of condensate water The water evaporates automaticall...

Page 23: ...d electronic appliances Before scrapping the machine make it unusable by cutting the power cord and removing the doors shelves and drawers so that children cannot access the unit Do not leave it unatt...

Page 24: ...e Manufacturer name Year of production Ambient temperature class Light power Defrost heater power Serial number Fuse Model Code Power voltage supply Power absorption Gas refrigerant type Ambient tempe...

Page 25: ...ers Labeling 21 Other Labeling Table 8 Labels Label Description Cut off power before removing the guard Periodic condenser cleaning Earthing symbol Attention risk of fire Do not exceed the indicated l...

Page 26: ...ezers Consumable Materials Table 9 Consumable Material Type Characteristics Application Image BATTERY 3V Lithium Coin Cell Battery type CR 1220 Clock battery on the board BATTERIES type AA rechargeabl...

Page 27: ...et a few at a time after the temperature has stabilised Material was placed in the freezer area at room temperature i e 25 Store only already frozen products Prolonged or too frequent door openings Re...

Page 28: ...t a few seconds and then re connect the plug Switch the refrigerator back on from the keyboard and once the temperature is displayed press OK so that the acoustic and visual alarms disappear Air condi...

Page 29: ...cal system or accidental disconnection of the plug POWER FAIL HT Check list alarms H 00 m00 Replace button battery on electronic board NO BATTERY Replace the battery BATTERY FAILURE Replace the batter...

Page 30: ...ed within 5 days of discovery and in any case within the maximum 6 months warranty period The customer will only be entitled to the repair or replacement of goods with the absolute exclusion of any di...

Page 31: ...formity with DIRECTIVE 2006 42 EC Declaration of conformity with DIRECTIVE 2014 30 EC Declaration of conformity with DIRECTIVE 2014 35 EC Declaration of conformity with DIRECTIVE 2011 65 EC RoHS Decla...

Page 32: ...personnel who performed it 2 Have the problems been verified according to the information stated in the use and maintenance manual YES No Fill out the following table Note Failure to submit a properl...

Page 33: ...to the status data menu with and scroll the items noting the values in the table below Table 11 Service status data for support Item Value U M KEYPAD ON n POWER ON n RESET h CONTROLLER h COMP h COMP C...

Page 34: ...thermofisher com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2012 19 EU in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE marcato col...

Page 35: ...This page is intentionally left blank...

Page 36: ...ew Zealand 64 9 980 6700 Nordic Baltic CIS countries 358 9 329 100 Russia 7 812 703 42 15 Spain Portugal 34 93 223 09 18 Switzerland 41 44 454 12 12 UK Ireland 44 870 609 9203 USA Canada 1 866 984 376...