General Purpose Refrigerator and Freezers
Safety | 5
Warning on the Refrigerant
Gas
These devices contain R290 as refrigerant
(Flammable Gas).
Ensure to follow the precautions given below:
-
To avoid damages to the refrigerant circuit, do not
use mechanical devices to accelerate defrost pro-
cess.
-
Position the appliance in an environment of suitable
dimensions in compliance with EN 378: the environ-
ment must have a volume of 1m
3
every 8 grams of
R290 refrigerant. The quantity of gas contained in
the circuit is declared on the silver data plate
attached inside the cabinet.
-
Do not use the equipment if damaged.
-
Make sure that air grids are always free to guarantee
a certain ventilation to the device.
-
If there is refrigerant leakage, avoid using open
flames, remove from the device flammable products
and ventilate immediately the environment.
-
Do not store potential explosive substances (for
example spray cans containing flammable gases)
inside the device.
Do not use electrical appliances inside the refrigerated
cabinet.
If any malfunction, disconnect the equipment from the
mains.
Extraordinary maintenance should be performed only by
qualified personnel.