29
6. Ensure the fluid used is 100% water. Heat the fluid to at least
30°C, then turn unit to full cool.
7. Begin timing as soon as the START temp (20.7°C) is passed.
8. Stop timing when the STOP temp (19.3°C) is passed.
9. Elapsed time must be one minute or less.
Suction Discharge Pressure and Speed Check
Air-Cooled Standard and High Temperature Units (All Pump Types)
1
Unit
Suction(psi)
Discharge(psi)
Speed Check(°C/Minute)
HX-100
75 - 83
245 - 265
1.3 - 1.6
HX-150
70 - 73
240 - 270
2.0 - 2.5
Water-Cooled Standard and High Temperature Units (All Pump Types)
2
Unit
Suction(psi)
Discharge(psi)
HX-100
86 - 92
180
Speed check times not
HX-150
65
175
established.
NOTE: For low temperature units please call NESLAB.