Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 28 of 84
c. Install the eyepieces if applicable otherwise continue with step 15d.
The
eyepiece with reticle should be inserted in the right receptacle.
d. Using the knobs on the sides of z stage assembly, turn the knobs
counter-clockwise (toward you) to raise the z stage platform all the way
up, until it stops.
e. Install the condenser.
i. Locate the threaded hole on the side of the condenser.
Eyepiece with reticle