21. If you
DO NOT
have an iN10MX with imaging option
or
an iZ10 option,
continue with step 26. If you have an imaging option
or
an iZ10 option,
continue with step 22.
22.
Remove the main cover of the iN10.
Using a short handled Phillips
screwdriver, remove the 5 screws from under the top front of the iN10 and
using a long handled screwdriver remove the two screws at the rear base of
the iN10 as shown below. Set the screws aside and lift the cover straight up
to remove it from the base. Set the cover aside.
23.
Remove the array detector shipping restraint if applicable and install
the Dewar fill extender (iN10MX with imaging option only). If an array is
not installed, continue with step 24.