Installing the iN10/iN10MX
This document contains confidential or proprietary information of Thermo Fisher Scientific. Neither this
document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or
in part, except as specifically authorized by Thermo Fisher Scientific.
8/3/2020
Page 37 of 61
42.1.1.
Run the transmission MCT performance test if applicable:
if an MCT is not installed, continue with step 42.2.
Click the MCT detector icon, and then click
Run Test.
When
the
Select Object and Final Format
window
appears, confirm
that the reflectance measurement box is NOT checked, click
OK to close the window
and then click OK to start the test.
Record the test results for the IQ documentation if applicable
and then continue with step 42.2.
42.2.
Prepare to run the system performance test in reflection mode.
From the
Window
drop down menu choose
Beampaths
and set
the
Collection Mode
to Reflection and the
Detector
to Single Point
MCT. If an MCT detector is not installed on the system, select
DTGS and continue with the following steps. Close the Beampath
window.