43.7.
Fine tune the first array detector mirror.
If the IR is shifted in any
direction between 5 and 25 microns from the video image, use a 5/64”
balldriver to adjust the appropriate screw on the first array detector
mirror to adjust the IR position. Refer to the illustrations below. Repeat
steps 43.5 through 43.7 until the IR image overlays the video image as
depicted in step 43.6 and then continue with step 44. If the IR image is
shifted
greater
than 25 microns in any direction; contact the factory.
IR position adjustment
up/down.
CW = Down
CCW = Up
IR position adjustment
left/right.
CW = Right
CCW = Left