Installing the iN10/iN10MX
This document contains confidential or proprietary information of Thermo Fisher Scientific. Neither this
document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or
in part, except as specifically authorized by Thermo Fisher Scientific.
8/3/2020
Page 49 of 61
44.
Verify the array zoom aperture alignment.
44.1.
From the View and Collect menu, turn the Imaging zoom optics
“on” by checking the
Imaging zoom optics
box as indicated
below.
44.2.
Collect a pinhole map.
Re-center the pinhole in the FOV if
necessary and shrink the area map size to reduce collection time,
center the map on the pinhole and collect the pinhole map.
44.3.
Analyze map.
If the IR and Video overlay as depicted on the following
page, proceed with step 45. If the IR and Video are shifted in any
direction, refer to the appropriate steps in the latest version of
Ser-004
Replacing the array foil zoom assembly
to re-align the zoom aperture
and then continue with step 45.