Installing the iN10/iN10MX
This document contains confidential or proprietary information of Thermo Fisher Scientific. Neither this
document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or
in part, except as specifically authorized by Thermo Fisher Scientific.
8/3/2020
Page 57 of 61
48.
Validate the iZ10 if applicable otherwise skip to step 51.
Close OMNIC
Picta and launch OMNIC. From the
Collect
drop down menu choose
Experiment Setup
and assure the
Sample Compartment
is identified as
iZ10 Right AEM and that a signal is present. Close the
Experiment Setup
window.
48.1.
Open Advanced Diagnostics.
From the
Collect
drop down menu
choose
Advanced Diagnostics
.