32
Countess
™
II Automated Cell Counter User Guide
Software update
Guidelines for
software update
•
The USB drive used for transferring the software update file must be FAT32
formatted; verify this before proceeding. If necessary, reformat the USB drive
to FAT32 following the recommended procedure for your operating system.
Note:
Reformatting the USB drive will result in the loss of all files. Back up the
files in the USB drive prior to reformatting.
•
The software update file must be saved on the top level of the USB drive, not
within a folder or a subfolder.
•
The software update file must be uncorrupted during transfer. Do not rename,
zip, or compress the software update file.
Update the
Countess
™
II/II FL
software
1.
Go to
www.thermofisher.com/countessupdate
, and download the latest
Countess
™
II/II FL software version to your desktop.
Note:
The software update file has a version-specific name followed by the
extension .lft (e.g., Countess_II_v_1_0_202.lft for software version 1.0.202).
2.
Copy the software update file onto the USB drive, making sure that it is saved
on the top level and not hidden within a folder.
3.
Insert the USB drive into one of the USB ports of the instrument (see page 8).
4.
Press the
Instrument Settings
button on the Home page (page 10) to
open the Instrument Settings screen (page 31).
5.
Select
Software Update
from the Instrument Settings menu. The instrument
scans the USB drive for the latest software version.
6.
When prompted, select
Update Now
.
7.
Once the update has completed, restart the instrument.