Operating the Microscope
Thermo Scientific
DXRxi User Guide
33
The transmitted light fixed polarizer fits in the transmission illumination assembly below the
stage platform.
Figure 34.
Transmitted light fixed polarizer insertion slot on the microscope
To install the transmitted light fixed polarizer
1. Open the sample compartment doors and remove any sample from the microscope stage.
2. Lower the stage just enough to clear the selected objective and then move the stage toward
the back of the microscope as far as it will go.
3.
Remove the sample area cover
.
4. If the handle is not attached to the fixed transmitted light polarizer, screw the handle into
the hole at the end of the polarizer and tighten it finger tight.
5. Hold the polarizer by the handle (label should face up) while you gently slide it into the
insertion slot under the stage platform.
Figure 35.
Transmitted light polarizer properly installed
6. Replace the sample area cover.
Insertion slot for fixed polarizer
for the transmitted light illuminator
Fixed transmitted light polarizer
properly installed
Summary of Contents for DXR xi
Page 12: ...1 Introducing the DXRxi Raman Imaging System Trademarks 6 DXRxi User Guide Thermo Scientific...
Page 14: ...2 DXRxi Microscope Operation 8 DXRxi User Guide Thermo Scientific...
Page 64: ...Safety Information 58 DXRxi User Guide Thermo Scientific Microscope Safety Labels or...
Page 65: ...Safety Information Thermo Scientific DXRxi User Guide 59...
Page 66: ...Safety Information 60 DXRxi User Guide Thermo Scientific...
Page 67: ...Safety Information Thermo Scientific DXRxi User Guide 61...
Page 68: ...Safety Information 62 DXRxi User Guide Thermo Scientific...
Page 166: ...3 OMNICxi Software Operation Acquire and Analyze Data 160 DXRxi User Guide Thermo Scientific...