Chemical Image Does Not Achieve Required Spatial Resolution
• Check if the correct objective is selected and in position for the spacial resolution selected
• Switch to video mode and make sure the sample is in focus
• Check the sample slide to make sure it is properly seated
• Check the aperture setting, make sure 25 micro pinhole was selected for high spatial
resolution imaging
• Lower the image pixel size setting in the software
• Switch to a shorter wavelength laser
Spatial resolution depends on the wavelength of the laser.
Chemical Image Shows Missing Data
• Increase the exposure time in the software
• Increase the number of scans in the software
• Reduce vibrations in the environment
• Use a vibration isolation table
Make sure that the isolation is active (for example, there is air pressure to the table)
X-Axis Calibration Fails
The neon bulb used for automatic x-axis calibration may have failed.
• Turn off automatic x-axis calibration and run manual calibration. Contact our service
representative in your area for repair of the automatic x-axis calibration feature.
Summary of Contents for DXR xi
Page 12: ...1 Introducing the DXRxi Raman Imaging System Trademarks 6 DXRxi User Guide Thermo Scientific...
Page 14: ...2 DXRxi Microscope Operation 8 DXRxi User Guide Thermo Scientific...
Page 64: ...Safety Information 58 DXRxi User Guide Thermo Scientific Microscope Safety Labels or...
Page 65: ...Safety Information Thermo Scientific DXRxi User Guide 59...
Page 66: ...Safety Information 60 DXRxi User Guide Thermo Scientific...
Page 67: ...Safety Information Thermo Scientific DXRxi User Guide 61...
Page 68: ...Safety Information 62 DXRxi User Guide Thermo Scientific...
Page 166: ...3 OMNICxi Software Operation Acquire and Analyze Data 160 DXRxi User Guide Thermo Scientific...