4
Troubleshooting
Error Messages
180
DXRxi User Guide
Thermo Scientific
Invalid fiber probe configuration
The installed fiber probe is not compatible with the current laser set (laser or grating). Install a
compatible laser set.
Uncalibrated laser, grating, filter set
The laser, grating, and filter set installed in the instrument has not been aligned and
calibrated. The instrument cannot collect data until the installed laser set (laser, filter, grating)
has been aligned and calibrated. For instructions, see
Aligning and Calibrating Laser Sets
.
Alignment/calibration out of date
The alignment and calibration for the laser, grating, and filter installed in the instrument
occurred longer ago than the specified interval. The instrument cannot collect data until the
installed laser set (laser, filter, grating) has been aligned and calibrated. For instructions, see
Aligning and Calibrating Laser Sets
Stage firmware out of date
A newer version of firmware for the stage controller is required.
Sample door open
The instrument cannot collect data because a sample compartment door is open or a laser
interlock override is missing (Class 3B instrument configurations only). Close both sample
compartment doors (or install the override), wait for the laser to warm up and restart data
collection.
Laser off
The instrument cannot collect data because the laser setting is off in the software. From the
OMNICxi Acquisition window, open the Settings box, select
Laser On
, close the Settings box
and restart data collection.
Laser key off
The instrument cannot collect data because the laser keyswitch is off. Set the laser keyswitch
to the On position and restart data collection.
Laser warming up
The instrument may not produce stable results right now because the laser is warming up.
Depending on the type of laser installed, allow 1 to 15 minutes for the laser to warm up and
then restart data collection.
Summary of Contents for DXR xi
Page 12: ...1 Introducing the DXRxi Raman Imaging System Trademarks 6 DXRxi User Guide Thermo Scientific...
Page 14: ...2 DXRxi Microscope Operation 8 DXRxi User Guide Thermo Scientific...
Page 64: ...Safety Information 58 DXRxi User Guide Thermo Scientific Microscope Safety Labels or...
Page 65: ...Safety Information Thermo Scientific DXRxi User Guide 59...
Page 66: ...Safety Information 60 DXRxi User Guide Thermo Scientific...
Page 67: ...Safety Information Thermo Scientific DXRxi User Guide 61...
Page 68: ...Safety Information 62 DXRxi User Guide Thermo Scientific...
Page 166: ...3 OMNICxi Software Operation Acquire and Analyze Data 160 DXRxi User Guide Thermo Scientific...