Operating the Microscope
46
DXRxi User Guide
Thermo Scientific
4. Connect the power cable for the probe LED.
Figure 53.
Connect power cable for probe LED
5. Attach the strain relief hook (provided) to the microscope by inserting the two screws
(provided) into the two holes at the top of the microscope back panel (behind the laser
filter compartment). Tighten the two screws.
6. Attach the probe cable to the strain relief hook.
Loop the nylon strain relief attached to the probe cable (before it splits) over the hook
installed in the previous step.
7. Open the
Settings
box in the OMNICxi Acquisition window and click
Align/Calibrate
and choose OK to align and calibrate the instrument with the probe.
The Fiber Launcher includes two caps to cover the ports when the accessory is not attached.
Always replace the caps to cover the ports when the accessory is not in connected.
Installing the Microscope Alignment and Calibration Tool
The alignment and calibration tool is connected to the instrument during installation, and
only needs to be disconnected if the sample area cover is removed by the user. The tool is
designed to fit the standard
that comes with the microscope.
Summary of Contents for DXR xi
Page 12: ...1 Introducing the DXRxi Raman Imaging System Trademarks 6 DXRxi User Guide Thermo Scientific...
Page 14: ...2 DXRxi Microscope Operation 8 DXRxi User Guide Thermo Scientific...
Page 64: ...Safety Information 58 DXRxi User Guide Thermo Scientific Microscope Safety Labels or...
Page 65: ...Safety Information Thermo Scientific DXRxi User Guide 59...
Page 66: ...Safety Information 60 DXRxi User Guide Thermo Scientific...
Page 67: ...Safety Information Thermo Scientific DXRxi User Guide 61...
Page 68: ...Safety Information 62 DXRxi User Guide Thermo Scientific...
Page 166: ...3 OMNICxi Software Operation Acquire and Analyze Data 160 DXRxi User Guide Thermo Scientific...