OdaLog
®
Gas Logger User Manual
P/N: 10-1000
Revision: F, Issued: October 2019
Page: 7
5.3 Null (zero)
With unit in fresh air press & hold button down until “NULL” is displayed and then release. Move magnet over
wording “NULL” on LHS of LCD screen – “DONE” should appear briefly when completed. (See Section 9.5)
5.4 Pre-Use Span Check
Apply a known level of gas over sensor (using supplied calibration fitting, [Part No. 11-0001]) at the flow rate
and for the flow time as detailed in Table 5 (or until reading stabilises) and record “Start Level” result (See
Section 9.8.1).
5.5 Start Logging (*Manual Logging setting only)
Press & hold button down until “START LOG” is displayed and then release (logging is confirmed by
LOGGING
symbol flashing on screen and flashing LED’s). (See Section 6.5)
5.6 Stop Logging (*Manual Logging setting only)
Press & hold button down until “STOP LOG” is displayed and then release. (See Section 6.6)
5.7 Post-Use Span Check
Flow a known level of gas over sensor (using supplied calibration fitting, [Part No. 11-0001]) at the flow rate
and for the flow time as detailed in Table 5 (or until reading stabilises) and record “End Level” result when
reading stabilises. (See Section 9.8.2)
5.8 Download Data
5.8.1 Use irDA
Stop logging as above then press & hold button down until “IrDA” is displayed then release. Use Infra Red Data
Communications link [Part No. 12-0002] and follow the
OdaStat-G™
software package instructions for
communicating with the OdaLog
®
instrument. (Refer to Section 6.8 for more information on communication
setup and download of data)
5.8.2 Use Bluetooth
Stop logging as above then press & hold button down until “BT” is displayed then release. Use Bluetooth Data
Communications link and follow the
OdaStat-G™
software package instructions for communicating with the
OdaLog
®
instrument.
5.9 Switch Off
Press & hold button down until “OFF” is displayed. Continue holding button as it counts down OFF-2-1 then
release. (See Section 6.10)
5.10 Training
To locate your nearest Thermo Fisher Scientific training centre please contact us at the address shown on the
back of this manual or refer to our web site
www.odalog.com