6.4 Displaying current status on the LCD
Press & hold button until “STATUS” is displayed then release button. The display will cycle through
“STATUS” screens as shown.
6.5 Start Logging Function (*Manual Logging Mode only, [Default =
Enabled])
Press & hold button down and continue holding as screen displays “STATUS’, then release button when
“START LOG” is displayed. See note 2 below if “CLOCK not SEt” is displayed when button released.
The OdaLog
®
is now logging as confirmed by the flashing “LOGGING” symbol on the display and the LED’s
on either side of the display flashing once every 10 seconds (Default: Enabled - can be changed in the
OdaStat-G™
software).
Note 1:
If the Logger memory is full and you have “Stop log when full” active, (
Refer to the “Instrument
Parameters” section in the
OdaStat-G™
software package
) the unit will not start logging, the display
will show “LOG” “FULL” and the LED’s will illuminate at the same time. The message will remain
on the display until the button is pressed to continue.
Note 2:
If the clock has not been set “CLOCK not SEt” will be displayed. The message will remain on the
display until the button is pressed to continue. (See Section 6.3)
6.6 Stop Logging Function (*Manual Logging Mode only, [Default =
Enabled])
Press & hold button down until “STOP LOG” is displayed then release button.