OdaLog
®
Gas Logger User Manual
P/N: 10-1000
Revision: F, Issued: October 2019
Page: 15
6.12 Low Battery Indication & Control
Battery power levels are indicated on-screen by means of the multi-segment battery symbol, as shown above
with the transitions with remaining life expectancy. When the battery is above 50% capacity all segments of the
will be displayed, as in stage 1 shown above. Similarly, stage 2 shown above indicates acceptable battery levels
down to 20% battery life.
At low voltage the battery symbol will be displayed as ‘empty’, i.e. without any segments visible, as shown at
stage 3 above. Then, as the battery capacity decreases to a flat (non-operational, stage 4) state, the OdaLog
®
instrument will stop the active logging session and switch the instrument off displaying the word “FLAT”
continuously on the display with the Battery Warning symbol flashing.
Note:
The instrument cannot be turned back on until the battery has been replaced. Refer Section 8.1.1 for
battery replacement.
6.13 Key Lock (Default = Disabled)
Key Lock (when enabled in the
OdaStat-G™
software): this feature enables the user to ‘lock’ the top button and
disable the “INST” display while in logging mode. This feature works as follows:
1.
The Key Lock option requires enabling from the
OdaStat-G™
software package, as per the “Instrument
Parameters” section.
2.
To Lock the OdaLog
®
instrument.
Once a logging session is started, the user can then select from the menu the option called “LOCK”.
With “LOCK” displayed, release the button to lock prompt confirmation. Swipe the magnetic base of
the Allen Tool provided past the right-hand side of the LCD display near the “Y” on screen. The LCD
will display “LOCKED”, as well as the “LOGGING” symbol flashing. The instrument and button
operation is now locked.