OdaLog
®
Gas Logger User Manual
P/N: 10-1000
Revision: F, Issued: October 2019
Page: 22
Remove filter
Remove, clean and replace stainless filter grill (if fitted)
Replace filter with ‘shiny’ side facing outwards
Replace o-ring. DO NOT GREASE O-RING
Replace filter housing assembly on OdaLog
®
body (aligning screw holes)
Replace three end cap-retaining screws. DO NOT OVERTIGHTEN
Note 1:
Filter orientation is important. Ensure that the replacement filter is correctly placed with the ‘shiny’
side facing outward.
Note 2:
Some sensor filters are fitted with stainless wire gauze protection. If fitted, this should be cleaned
and dried prior to fitting of new filter.
8.2 Service
To locate your nearest Thermo Fisher Scientific service centre please contact us or refer to our web site:
www.odalog.com
NOTE:
Ensure hands are free of dirt and grease prior to handling new filter