OdaLog
®
Gas Logger User Manual
P/N: 10-1000
Revision: F, Issued: October 2019
Page: 27
Ensure that instantaneous display reads the same as shown on cylinder before turning off gas regulator and
removing calibration fitting.
Ensure that reading returns to close to zero within 5 minutes in clean air and re-zero if necessary.
Note:
The instrument will TIMEOUT and return to the ‘INST’ screen if a SPAN is not done within 10min.
9.7 Calibration Errors
Note: If ‘FAILED”, “SENSOR bAd1” or “SENSOR bAd2” appears on screen during SPAN operation,